Remote Control - Commands
NRP2
336User Manual 1173.9140.02 ─ 07
UNIT<[1]..8>:POWer <output_unit>
Selects the output unit for the measured power values.
This setting also determines the unit for the parameters of the following commands:
CALC:REL
CALC:LIM:UPP
CALC:LIM:
LOW
DISP:MET:UPP
DISP:MET:LOW
OUTP:REC:LIM:LOW
OUTP:REC:LIM:UPP
Parameters:
<output_unit> DBM | W | DBUV
*RST: DBM
UNIT<[1]..8>:POWer:RATio <output_unit>
Selects the output unit for the measured power ratio values.
This setting also determines the unit for the parameters of the following commands:
CALC:REL
CALC:LIM:UPP
CALC:LIM:
LOW
DISP:MET:UPP
DISP:MET:LOW
OUTP:REC:LIM:LOW
OUTP:REC:LIM:UPP
Error messages:
If the compute function of the associated calculate block is :SWR, :RLOSs or :
REFLection, then the unit cannot be changed and error message 26,"State not sup-
ported", is output.
Parameters:
<output_unit> DB | DPCT | O
*RST: DB
UNIT