Getting Started
NRP2
31User Manual 1173.9140.02 ─ 07
Setting the unit for the measurement result
The measurement result is indicated in the result window. The unit of the measurement
result can be changed.
To set the unit of the measurement result
1. Use the topmost softkey to select the "Measurement" menu.
The menu opens. Post-processing of the measurement can be set in the "Mea-
surement" menu.
Fig. 2-5: Measurement menu
2. Select the unit "W" using the associated softkey.
The result is displayed in W.
Fig. 2-6: Result window (W)
Setting a fixed offset correction
The R&S NRP2 provides an offset correction to the displayed measurement result.
To set a fixed offset correction
1. Select the "Sensor" menu.
The "Sensor" menu opens.
Measuring average power (Cont Av mode)