Error Messages and Troubleshooting
NRP2
342User Manual 1173.9140.02 ─ 07
Error code Error text returned upon queue query
Explanations
-168 "Block data not allowed"
The command contains a legal block data element at a point where it is not allowed.
-171 "Invalid expression"
The command contains an invalid mathematical expression data element.
-178 "Expression data not allowed"
The command contains mathematical expression data at a point where they are not allowed.
8.2.3 Execution Error
Execution errors; cause bit 4 in the ESR register to be set
Error code Error text returned upon queue query
Explanations
-200 "Execution error"
An error occurred upon command execution.
-203 "Command protected"
An attempt was made to execute a protected command.
-210 "Trigger error"
Error on triggering the device.
-211 "Trigger ignored"
A *TRG or a triggering signal was ignored.
-213 "Init ignored"
A request for a measurement initialization was ignored as another measurement was already
in progress.
-214 "Trigger deadlock"
A measurement could not be started or a result query could not be processed since the
device would otherwise enter a deadlock state. This occurs under the following conditions:
(1)TRIG:SOUR BUS;:
INIT;:FETCh?: A *TRG or TRIG:IMM would be required for trigger-
ing but these commands cannot be executed since FETCh? blocks the entry. But FETCh?
waits until the measured values are available.
(2) FETCh? was called but no valid result was available and the trigger system of the sensor
is IDLE. FETCh? blocks the entry so that a measurement cannot be started with INIT
either.
-220 "Parameter error"
The command contains a faulty or invalid parameter.
-221 "Settings conflict"
There is a setting conflict between two or more parameters.
-222 "Data out of range"
A program data element is outside the legal range as defined by the device.
SCPI-Specific Error Messages