Error Messages and Troubleshooting
NRP2
343User Manual 1173.9140.02 ─ 07
Error code Error text returned upon queue query
Explanations
-223 "Too much data"
A valid parameter is received but it contains more data than the device can handle.
-224 "Illegal parameter value"
The parameter value is invalid.
-225 Out of memory
The device has insufficient memory to perform the required operation.
-226 "Lists not same length"
A list of offset values and frequencies has not the same length as an associated list of fre-
quency and offset values.
-230 "Data corrupt or stale"
The data are incomplete or invalid.
-240 "Hardware error"
The command cannot be executed because of a hardware problem in the device.
-241 "Hardware missing"
The command cannot be executed because of missing device hardware.
8.2.4 Device-Specific Errors
Device-specific error; causes bit 5 to be set in the ESR register
Error code Error text returned upon queue query
Explanations
-300 "Device-specific error"
Generic device-dependent error that cannot be defined more precisely.
-310 "System error"
Indicates that some system error has occurred. Please contact the R&S service center.
-311 "Memory error"
An error was detected in the device's memory.
-314 "Save/recall memory lost"
Indicates that the nonvolatile data saved by the *SAV? command has been lost.
-315 "Configuration memory lost"
Indicates that nonvolatile configuration data saved by the device has been lost.
-330 "Self-test failed"
The self-test could not be executed.
-341 "Zeroing failed"
The zeroing of a sensor (CALibration:ZERO:
AUTO ONCE) has failed. The most frequent
cause for this error is a signal that is applied to the sensor input and prevents zeroing.
SCPI-Specific Error Messages