Error Messages and Troubleshooting
NRP2
345User Manual 1173.9140.02 ─ 07
Error code Error text returned upon queue query
Explanations
24 "Sensor mode not supported"
An attempt was made to set the sensor to a measurement mode which is not supported.
25 "Not supported"
A command is not supported (by the sensor).
26 "State not supported"
Attempt was made to set an invalid value. The permissible values for sensor-dependent
commands can be declared as impermissible by a sensor depending on the context.
27 "Sensor not present"
A command could not be executed because the required sensor is not connected.
28 "Sensor not idle"
A command could not be executed because the respective sensor was not in the IDLE state
at that time.
29 "Measurement aborted while waiting for data"
A measurement was aborted while the device was waiting for measurement results.
30 "No extremes available in this mode"
Extreme values (minimum, maximum und peak-to-peak values) are not calculated in the set
measurement mode.
39 "Sensor removed while waiting for result"
A sensor was disconnected while the device was waiting for measurement results.
42 "Failed to initialize sensor"
A sensor could not be initialized.
43 "Error in receiving calibration data from sensor"
An error occurred in receiving calibration data from the sensor (command
CALibration[1..4]:
DATA?).
44 "Error in sending calibration data to sensor"
An error occurred in sending calibration data to the sensor
(command
CALibration[1..4]:DATA <block_data>).
45 "Command not supported by sensor"
Attempt was made to send a command to a sensor that does not support this command.
This may depend on the current sensor status. The sensors can inform the R&S
NRP2 at
any point which commands they support or do not support. This mainly concerns commands
of the Sense and Trigger systems.
46 "Sensor failure: no command receipt"
A sensor did not acknowledge any command.
47 "Error in receiving battery data"
An error occurred in receiving battery data (command SYSTem:BATTery:
STATus?).
48 "Error in sending battery data"
An error occurred in sending battery data (command SYSTem:BATTery:MODE
<block_data>).