EasyManua.ls Logo

Rohde & Schwarz NRP2 - Page 400

Rohde & Schwarz NRP2
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Annex
NRP2
399User Manual 1173.9140.02 ─ 07
The following example is intended for analysing a GSM/EDGE signal with the first slot
used.
*rst
*cls
sens:trac:meas on 'activate pulse measurements
sens:trac:meas:auto on 'for sensor A.
sense:trace:meas:tran:esam:auto:state on 'enhanced accuracy for stable
'periodic signals
sens:trac:meas:alg HIST 'select an algorithm for the pulse
'detection analysis
meas:xtime? (100), 6 ms, (@1) 'perform a trace measurement
'Check the reference levels for
sens:trac:meas:def:dur:ref? '- pulse duration
sens:trac:meas:def:tran:lref? '- start of rising/end of falling edge
sens:trac:meas:def:tran:href? '- start / end of pulse top
sens:trac:meas:puls:dcyc? 'read duty cycle
sens:trac:meas:puls:dur? 'read pulse duration
sens:trac:meas:puls:per? 'read pulse period
sens:trac:meas:puls:sep? 'read pulse separation
sens:trac:meas:tran:neg:dur? 'read length of falling edge
sens:trac:meas:tran:neg:occ? 'read start time of falling edge
sens:trac:meas:tran:neg:over? 'read overshooting of falling edge
sens:trac:meas:tran:pos:dur? 'read length of rising edge
sens:trac:meas:tran:pos:occ? 'read start time of rising edge
sens:trac:meas:tran:pos:over? 'read overshooting of rising edge
sens:trac:meas:power:min? 'read minimum power in trace
sens:trac:meas:power:max? 'read maximum power in trace
sens:trac:meas:power:avg? 'read average power in trace
sens:trac:meas:power:puls:top? 'read mean power level of pulse top
sens:trac:meas:power:puls:base? 'read mean power level of pulse bottom
syst:err? 'check for errors
A.4.7 Synchronizing Measurements
Some sensors (e.g. R&S
NRP-Z81/85/86) are able to generate signals on the (exter-
nal) trigger bus, when an internal trigger was detected. This allows for synchronized
measurements with multiple sensors.
One sensor has to be configured as trigger master (:TRIG:SOUR INT; :TRIG:MAST
ON), while all other sensors are slaves with external trigger source (:TRIG:SOUR
EXT).
Programming Measurements
北京海洋兴业科技股份有限公司(证券代码:839145)