Instrument Functions
NRP2
80User Manual 1173.9140.02 ─ 07
In manual mode, “longer measurement” does not mean that it takes longer to display a
new result, but rather that it takes longer for the result to settle when the power
changes.
The filter settings of the "Trace & Statistics" mode and the other modes, which exclu-
sively provide average power values, differ significantly. The filter dialog box thus dis-
plays different parameters depending on the selected display mode. Filtering in the
"Trace & Statistics" mode is described in chapter 4.5.3.3, "Filtering settings",
on page
118.
All average power values are obtained from a three-stage averaging process. First, a
mean value is formed for the acquisition interval given in table 4-2
- either by weighted
or unweighted summation of samples.
A second measurement with reversed polarity of the detector output signal is then per-
formed to suppress offset voltages and 1/f noise of the amplifier. A first result is formed
by subtracting the second measured value from the first measured value. The
sequence of two consecutive measurements is repeated N times. The average value,
which is obtained from the individual measurement results, is output. The number N is
the averaging number ("AvgNo"). To obtain a high output data rate, averaging in man-
ual mode is performed continuously, i.e. a new result is fed into the averaging filter
after two acquisitions have been evaluated. The oldest result is taken out of the aver-
aging filter and a new average value is calculated.
The averaging number can be selected automatically or it can be manually set to a
fixed value. As a starting point, automatic filtering should always be used. If the auto-
matically selected filter setting proves to be not adequate, the averaging number can
be in(de)creased manually.
Two autofilter operating modes are provided:
"Normal"
The "Normal" mode finds a balance between measurement time and display noise.
As an additional parameter, display resolution can be used to adjust the filter set-
ting. With increasing resolution, i.e. an increasing number of digits in the measure-
ment results, the averaging number also increases.
"Fixed Noise"
In the "Fixed Noise" mode, the averaging number is selected so that the sensors
intrinsic noise (2 standard deviations) does not exceed the specified "Noise Con-
tent". To avoid very long settling times when the power is low, the averaging num-
ber can be limited with the "Max Settling Time" parameter. If the display noise then
exceeds the set noise content, the symbol "S/N" is displayed.