R&S PR100 User Manual
Remark:
All M-trace, I-trace and IFPan-trace buffers are cleared after executing an
INITiate command. Only the first measurement value is stored in the trace-
buffers when the measurement mode is set to continuous. Values will be
added until the trace-buffer is full when the mode equals periodic
Example:
INITiate
INITiate
. :CONM
. . [:IMMediate]
The INITiate CONtinue Measurement is identical to the INITiate command
described in the previous section except for the fact that trace buffers are not
cleared and an active scan is not restarted. Using the CONTinue
Measurement form appends new measurement values to the current values
present in the trace buffers. Using this command when a scanner causes the
scanner to select the next frequency in frequency scan mode or the next
memory channel when the memory scanner is active.
Parameters:
None
Remark:
The INITate CONtinue Measurement command has no effect when the
measurement mode is set to periodic.
Example:
INITiate:CONM