R&S PR100 User Manual
[SENSe]
. :MSCan
. . :CONTrol
. . . :[ON] <control_function> {,<control_function>}
Command for switch-on of the 'STOP:SIGNal' function. When this function is
off, the memory scan stops at each location with a signal for the dwell-time.
When this function is on, the dwell-time is controlled by the presence of a
signal stronger than the threshold level:
During a memory-scan, the receiver moves from one memory location to
another, loading the settings into the receiver. If a memory location has a
signal that is stronger than the threshold level, the receiver stays on that
memory location for a certain time, called dwell time. When the signal
disappears during that dwell time, the receiver stays on the same location for
a while, called hold time, to see if the signal re-appears. When either the hold-
time or the dwell-time has elapsed, scanning continues.
If the signal does re-appear, the receiver continues the dwell time again (the
dwell-time never stopped), otherwise it moves to the next memory location.
This value is mapped onto the same variable as the frequency scan control
mechanism variable, meaning that a change affects both scan modes.
Parameters:
<control_function> is the following string:
“STOP:SIGNal” switches the signal-controlled dwell time on
*RST state:
after *RST, all control mechanisms are on
Example:
MSCan:CONTrol “STOP:SIGN”