R&S PR100 User Manual
0,"No error" This message is output if the error queue does
Table 12: Command Error - Faulty command; sets bit 5 in the ESR
register
Error code from queue
query
The command contains an invalid sign.
Example: A command contains an ampersand,
"SENSe &".
-
The command is invalid. Example: The
command contains block data the instrument
does not accept.
-
The command contains an impermissible sign
instead of a separator. Example: A semicolon is
missing after the command.
-
The command contains an invalid value
indication. Example: ON is indicated instead of a
numeric value for frequency setting.
-
A Group Execute Trigger (GET) is within a
-108,"Parameter not
allowed"
The command contains too many parameters.
Example: Command SENSe:FREQuency
permits only one frequency indication.
-109,"Missing parameter" The command contains too few parameters.
Example: Command SENSe:FREQuency
requires a frequency indication.
-111,"Header separator
error"
The command contains an impermissible
separator. Example: The header is not followed
by a "White Space", "*ESE255"
-112,"Program
mnemonic too long"
The command contains more than 12
characters.
-
The command is not defined for the instrument.
Example *XYZ is undefined for every
instrument.
-114,"Header suffix out
of range"
The command contains an impermissible
numeric suffix. Example: SENSe9 does not exist
in the instrument.
-121,"Invalid character in
number"
A number contains an impermissible character.
-123,"Exponent too
large"
The absolute value of the exponent is larger
than 32000.
-
The number contains too many digits.