45Getting Started 1179.4657.02 ─ 10
Main power switch function:
Position 1: The instrument is in operation.
Position O: The entire instrument is disconnected from the AC power supply.
For details, refer to "Connecting to power" on page 9 and Chapter 4.5, "Connect-
ing to Power", on page 21.
5.2.2 Li-Ion Battery Packs and DC Power Connector
With the Li-ion battery pack (option R&S FPL1-B31), the R&S FPL1000 can be
operated independently of an AC or DC power supply. The instrument can house
2 Li-ion batteries which can be charged both via AC or DC power supply.
For safety information concerning batteries, see "Handling batteries safely"
on page 10.
As an alternative, a DC power supply connector (option R&S FPL1-B30) is availa-
ble. DC power supplies from +12 V to +24 V and from 13 A to 6.5 A can be used.
Connect the connector according to the following diagram:
Pin Description
1 2
3
1 Plus
2 Ground
3 Not used
If you use an external power supply unit to supply safety extra-low DC voltage
(SELV) to the instrument, be sure to meet the requirements for reinforced/double
insulation in accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No.
1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950). Provide current
limitation in accordance with DIN EN 61010-1 Appendix F2.1.
Also see "Connecting to power" on page 9.
5.2.3 GPIB Interface
The optional GPIB interface (R&S FPL1-B10) is in compliance with IEEE488 and
SCPI. A computer for remote control can be connected via this interface. To set
up the connection, a shielded cable is recommended.
Rear Panel View