R&S FSL CALCulate Subsystem (SPECM, K14)
1300.2519.12 6.391 E-11
C
ALCulate<1|2>:DELTamarker<1...4>:SPECtrogram:FRAMe
This command defines the frame number of the specified delta marker.
T
he numeric suffixes <1|2> are irrelevant for this command.
Parameter
0 to <number of recorded frames>
If the maximum number of frames is recorded, the max. value depends on the history depth (for
details refer to CALCulate<1|2>:SPECtrogram:HDEPth command)
Example
CALC:DELT3 ON
Switches on delta marker 3 or switches to delta marker mode.
CALC:DELT3:SPEC:FRAM –50
Positions delta marker 3 on frame number –50.
Characteristics
RST value: 0
SCPI: device–specific
Mode
SPECM
CALCulate<1|2>:DELTamarker<1...4>:SPECtrogram:SARea
This command defines the search area of the specified delta marker.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
VISible | MEMory
Example
CALC:DELT2 ON
Switches on delta marker 2 or switches to delta marker mode.
CALC:DELT2:SPEC:SAR MEM
Searches the whole data range stored in the history buffer.
Characteristics
RST value: VIS
SCPI: device–specific
Mode
SPECM
CALCulate<1|2>:DELTamarker<1...4>:SPECtrogram:XY:MAXimum[:PEAK]
This command positions the specified delta marker on the current maximum of the search area.
The search area is defined via the
CALCulate<1|2>:DELTamarker<1...4>:SPECtrogram:SARea command.
This command is an event and therefore has no *RST value and no query.
The numeric suffixes <1|2> are irrelevant for this command.