R&S FSL CALCulate Subsystem (CATV, K20)
1300.2519.12 6.451 E-11
CALCulate:DTV:LIMit:RESult:CFOFfset:UPPer?
This command performs an upper limit check for the carrier frequency offset in the Overview
measurement.
This command is a query and therefore has no *RST value.
Example
CALC:DTV:LIM:CFOF:UPP 20KHZ
Sets the upper limit for the carrier frequency offset to 20 kHz.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:DTV:LIM:RES:CFOF:UPP?
Performs an upper limit check for the carrier frequency offset.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate:DTV:LIMit:RESult:CHPower?
This command queries the result of the Channel Power measurement. Upper and lower limit
check can not be handled separately.
Example
CALC:DTV:LIM:CHP:UPP –18
Sets the upper limit for the Channel Power measurement to –18 dBm.
CALC:DTV:LIM:CHP:LOW –22
Sets the lower limit for the Channel Power measurement to –22 dBm.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:DTV:LIM:RES:CHP?
Performs a limit check for the Channel Power measurement.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate:DTV:LIMit:RESult:CSUPpression[:LOWer]?
This command performs a lower limit check for the lower limit of the carrier suppression in the
Modulation Errors measurement.
This command is a query and therefore has no *RST value.
Example
CALC:DTV:LIM:CSUP 30
Sets the lower limit for the carrier suppression to 30 dB.
INIT;*WAI
Starts a new measurement and waits for the sweep end.