R&S FSL DISPlay Subsystem (CATV, K20)
1300.2519.12 6.475 E-11
Example
CONF:DTV:MEAS CONS
Configures the R&S FSL for the Constellation Diagram measurement.
DISP:TRAC1:MODE:FRE ON
Sets the display mode of the constallation diagram to on.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
CATV
DISPlay[:WINDow<1|2>]:TRACe<1...4>:MODE:HCONtinuous
The numeric suffixes <1...4> select the trace.
For further details refer to "DISPlay[:WINDow<1|2>]:TRACe<1...6>:MODE:HCONtinuous" on
page 6.133..
DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y[:SCALe]
The numeric suffixes <1...4> select the trace.
For details refer to "DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]" on page 6.134.
DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y[:SCALe]:AUTO
This command activates the automatic scaling for the Hum and Tilt measurement. If activated,
the scaling is automatically optimized to the best display of data. This rescaling may affect the
values of any of the parameters under the SCALe node.
The numeric suffixes <1|2> and <1...4> are irrelevant.
This command is an event and therefore has no *RST value and no query.
Example
DISP:TRAC:Y:AUTO
Activates the automatic scaling for the Hum measurement.
Characteristics
RST value:
SCPI: conform
Mode
CATV
DISPlay[:WINDow<1|2>]:TRACe<1...4>:Y[:SCALe]:MODE
The numeric suffixes <1...4> select the trace.
For further details refer to "DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:MODE" on page
6.135.

Table of Contents