R&S FSL DISPlay Subsystem (CATV, K20)
1300.2519.12 6.477 E-11
D
ISPlay[:WINDow<1|2>]:ZOOM:EPATtern:STATe
This command activates or deactivates the zoom on around 0 Us for the Echo Pattern
measurement.
The numeric suffixes <1|2> are irrelevant.
Parameter
ON | OFF
Example
CONF:DTV:MEAS EPAT
Configures the R&S FSL for the Echo Pattern measurement.
DISP:ZOOM:EPAT:STAT ON
Switches on the zoom for the Echo Pattern measurement.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
CATV
DISPlay[:WINDow<1|2>]:ZOOM:MERRors
This command sets the zoom for the Modulation Errors measurement.
The numeric suffixes <1|2> are irrelevant.
Parameter
NONE The whole result table is displayed.
<parameter> The result table of the selected parameter is displayed.
with parameter = IMBalance | QERRor | SUPpression | PJITter | MERRms
| MERPeak | EVMRms | EVMPeak
Example
CONF:DTV:MEAS MERR
Configures the R&S FSL for the Modulation Errors measurement.
DISP:ZOOM:MERR MERP
Displays the measurement results of the peak of the modulation error rate.
Characteristics
RST value: NONE
SCPI: device–specific
Mode
CATV