R&S FSL CALCulate:MARKer Subsystem (WLAN, K91 / K91n)
1300.2519.12 6.691 E-11
CALCulate:MARKer Subsystem (WLAN, K91 / K91n)
The CALCulate:MARKer subsystem checks the marker functions of the instrument.
The following subsystem is included:
• "CALCulate:MARKer:FUNCtion Subsystem (WLAN, K91 / K91n)" on page 6.696
Commands of the CALCulate:MARKer Subsystem
– CALCulate<1|2>:MARKer<1>[:STATe]
– CALCulate<1|2>:MARKer<1>:AOFF
– CALCulate<1|2>:MARKer<1>:BSYMbol
– CALCulate<1|2>:MARKer<1>:CARRier
– CALCulate<1|2>:MARKer<1>:MAXimum
– CALCulate<1|2>:MARKer<1>:MINimum
– CALCulate<1|2>:MARKer<1>:SYMBol
– CALCulate<1|2>:MARKer<1>:TRACe
– CALCulate<1|2>:MARKer<1>:X
– CALCulate<1|2>:MARKer<1>:Y
CALCulate<1|2>:MARKer<1>[:STATe]
This command switches the markers on or off.
Parameter
ON | OFF
Example
CALC1:MARK1:STATE ON
Switches the screen A marker ON.
CALC2:MARK1:STATE OFF
Switches the screen B marker OFF.
Characteristics
RST value: ON
SCPI: device–specific
Mode
WLAN
CALCulate<1|2>:MARKer<1>:AOFF
This command switches off all active markers in the specified measurement window. The
window is set by the numeric suffix of CALCulate<1|2>.
Example
CALC1:MARK:AOFF
Switches off all markers in the screen A window.
CALC2:MARK:AOFF
Switches off all markers in the screen B window.