WLAN TX Measurements (Option K91 / K91n) R&S FSL
1300.2519.12 4.418 E-11
Capture Time (General Settings dialog box)
Specifies the time (and therefore the amount of data) to be captured in a single measurement
sweep.
Remote: SWE:TIME 10ms
Burst Count (General Settings dialog box)
Activates or deactivates a specified number of bursts for capture and analysis.
On The data analysis is performed over a number of consecutive sweeps until the
required number of bursts has been captured and analyzed.
Off The data analysis is performed on a single measurement sweep.
Remote: BURS:COUN:STAT ON
Analyze Bursts (General Settings dialog box)
Specifies the number of bursts to be measured, if the Burst Count option is activated.
Remote: BURS:COUN 16
Sweep Count (General Settings dialog box)
Specifies the number of sweeps to be performed for Spectrum ACP/ACPR and Spectrum Mask
measurements.
Remote: SWEep:COUNt 64
Trigger Mode (General Settings dialog box)
Sets the source of the trigger for the measurement sweep.
Free Run
The measurement sweep starts immediately.
External
Triggering via a TTL signal at the input connector EXT TRIGGER/GATE
IN on the rear panel.
Power
The measurement sweep starts when the signal power meets or exceeds
the specified power trigger level. This trigger mode is not available for
Spectrum Mask measurements in ETSI standard. If it is set and then the
Spectrum Mask measurement in ETSI standard is selected, it automatically
changes to Free Run.
Remote: TRIG:MODE IMM
Trigger Offset (General Settings dialog box)
Specifies the time offset between the trigger signal and the start of the sweep. A negative value
indicates a pre–trigger. This field is not available in the Free Run trigger mode.
Remote: TRIG:HOLD 500us