WLAN TX Measurements (Option K91 / K91n) R&S FSL
1300.2519.12 4.424 E-11
Level (Demod Settings dialog box)
Activates or deactivates the compensation for level error. If activated, the measurement results
are compensated for level error on a per–symbol basis.
Remote: TRAC:LEV ON
Display List/Graph
Configures the result display. The measurement results are displayed either in form of a list of
measurement points or as a graphical trace.
Remote: DISP:WIND1:TABL ON
Remote: for result queries see chapter "Remote Control – Commands", section FETCh Subsystem
PVT
Opens the PVT submenu to select the Power vs Time measurement results.
The PVT result displays show the minimum, average and maximum levels measured over the
full range of the measured input data, or over complete bursts displayed within the gating lines if
gating is switched on. The results are displayed as a single burst. Using screen B in full screen
provides additional power information during this measurement.
For IEEE 802.11b and g (single carrier), the PVT results are displayed as percentage values of
the reference power. The reference can be set to either the max or mean power of the burst. For
both rising and falling edges two time lines are displayed, which mark the points 10% and 90%
of the reference power. The time between these two points is compared against the limits
specified for the rising and falling edges.
For further details see also "Rise / fall time measurement" on page 4.413.
Remote: CONF:BURS:PVT