WLAN TX Measurements (Option K91 / K91n) R&S FSL
1300.2519.12 4.426 E-11
Ramp Up/Down/Up & Down (IEEE 802.11b, g – Single Carrier)
Sets the display of the rising/falling edge graph:
Up
Displays the rising edge graph.
Down
Displays the falling edge graph.
Up & Down
Displays the rising and falling edge graph.
For further details refer to the PVT softkey.
Remote: CONF:BURS:PVT:SEL RISE
Ref Pow Max/Mean (IEEE 802.11b, g – Single Carrier)
Sets the reference for the rise and fall time calculation to the maximum or mean burst power.
For further details refer to the PVT softkey.
Remote: CONF:BURS:PVT:RPOW MEAN
Average Length (IEEE 802.11b, g – Single Carrier)
Opens an edit dialog box to enter the number of samples in order to adjust the length of the
smoothing filter.
For further details refer to the PVT softkey.
Remote: CONF:BURS:PVT:AVER 31
Gating Settings On/Off
Activates or deactivates gating, or opens the Gate Settings dialog box to specify range of
captured data used in results calculation.
On Uses only the specified range of captured data in results calculation. In the
Magnitude Capture Buffer trace, two vertical lines mark the specified range.
Off Uses all the captured data in results calculation.