EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 612

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL WLAN TX Measurements (Option K91 / K91n)
1300.2519.12 4.427 E-11
In the Gate Settings dialog box, the following parameters are set:
Delay
Start point of captured data to be used in results calculation, i.e.
the delay from the start of the captured data in time or samples. If
the delay is specified in time, the number of samples is updated
a
ccordingly, and vice versa.
Length
Amount of captured data to be used in results calculation. If the
length is specified in time, the number of samples is updated
accordingly, and vice versa.
Link Gate and Mark
If activated, the position of the marker and the gate lines are
linked. The marker is positioned half way between gate start and
end. The marker position alters when the gate is modified, and
the gate lines move with the marker when the marker position is
altered.
The gate settings are defined for following measurements: PVT, Spectrum FFT, CCDF,
Spectrum Mask, Spectrum ACPR.
If a frequency sweep measurement is active (Spectrum Mask and Spectrum ACP) the result
display is switched to the Magnitude Capture Buffer display in order to allow the gate to be set
the correct part of the sweep.
Remote: SWE:EGAT ON
Remote: SWE:EGAT:HOLD 125us, SWE:EGAT:HOLD:SAMP 2500 (Delay)
Remote: SWE:EGAT:LENG 20ms, SWE:EGAT:LENG:SAMP 200000 (Length)
Remote: SWE:EGAT:LINK ON (Link Gate and Mark)
Import
Opens the Choose the file to import dialog box.
Select the IQ data file you want to import and press ENTER. The extension of data files is *.iqw.
Remote: MMEM:LOAD:IQ:STAT
Export
Opens the Choose the file to export dialog box.
Enter the path and the name of the IQ data file you want to export and press ENTER. The
extension of data files is *.iqw. If the file cannot be created or there is no valid IQ data to export
an error message is displayed.
Remote: MMEM:STOR:IQ:STAT
R&S Support
Creates files that help identifying the current problem. They are stored under
C:\R_S\Instr\user\Support. If you contact the Rohde&Schwarz support to get help for a certain
problem, send these files to the support in order to identify and solve the problem faster.