WLAN TX Measurements (Option K91 / K91n) R&S FSL
1300.2519.12 4.436 E-11
PLCP Header
This result display shows the decoded data from the PLCP header of the burst. The following
details are listed:
Column header Description Example
Burst number of the decoded burst
A colored block indicates that the burst was successfully
decoded.
Burst 1
Signal signal field
The decoded data rate is shown below.
00010100
2 MBits/s
Service service field
The currently used bits are highlighted. The text below explains
the decoded meaning of these bits.
00000000
––/––/––
PSDU Length length field
The decoded time to transmit the PSDU is shown below.
0000000001111000
120 Ms
CRC CRC field
The result is displayed below (OK for passed or Failed).
1110100111001110
OK
Remote: CONF:BURS:STAT:SFIeld
Softkeys of the sweep menu (WLAN mode)
The following table shows all softkeys available in the sweep menu in WLAN mode (SWEEP key). It is
possible that your instrument configuration does not provide all softkeys. If a softkey is only available
with a special option, model or (measurement) mode, this information is delivered in the corresponding
softkey description.