R&S FSL WiMAX, WiBro Measurements (Options K92/K93)
1300.2519.12 4.455 E-11
Sweep Time (General Settings dialog box)
Specifies the sweep time for the Spectrum Mask and Spectrum ACP/ACPR measurements:
–
If the Auto option is activated, the sweep time is calculated automatically.
– If the Auto option is deactivated, specify the sweep time.
Remote: SWE:ACPR:TIME 1S
Remote: SWE:ACPR:TIME:AUTO OFF
Sweep Count (General Settings dialog box)
Specifies the number of sweeps to be performed for Spectrum ACP/ACPR and Spectrum Mask
measurements.
Remote: SWE:COUNt 64
Trigger Mode (General Settings dialog box)
Sets the trigger source for the measurement sweep.
Free Run
The measurement sweep starts immediately.
External
The measurement sweep starts if the external trigger signal meets or
exceeds the external trigger level (a fixed value that cannot be altered) at
the input connector EXT TRIGGER/GATE IN on the rear panel.
Power
The measurement sweep starts if the signal power meets or exceeds the
specified power trigger level.
Remote: TRIG:MODE IMM
Trigger Offset (General Settings dialog box)
Specifies the time offset between the trigger signal and the start of the sweep. A negative value
indicates a pre–trigger. This field is not available in the Free Run trigger mode.
Remote: TRIG:HOLD 500us
Power Level (General Settings dialog box)
Specifies the trigger level if the Power trigger mode is set (see Trigger Mode field).
Remote: TRIG:LEV:POW 10 DBM
Auto Lvl (General Settings dialog box)
Activates or deactivates the automatic measurement of the power trigger level if the Power
trigger mode is set (see Trigger Mode field).
On The power trigger level is measured automatically at the start of each measurement
sweep. This ensures that the power trigger level is always set at the optimal level for
obtaining accurate results but will result in a slightly increased measurement times.
Off
The power trigger level is defined manually in the Power Level field.
Remote: TRIG:LEV:POW:AUTO ON