R&S FSL Instrument Setup and Interface Configuration - SETUP Key
1300.2519.12 4.495 E-11
Transducer
Opens the Select Transducer dialog box and a submenu to activate or deactivate defined
transducer factors, to generate new transducer factors or to edit existing ones.
In the Select Transducer dialog box, a transducer factor can be edited, deleted or copied, if
selected, using the corresponding softkeys. The default directory, in which the transducer factors
are saved, is displayed (C:\R_S\INSTR\TRD; the extension of transducer factor files is *.TDF).
Subdirectories can be added or deleted via the file manager. They are displayed or hidden using
the arrow keys or the Show Directory softkey / Hide Directory button.
Two different directory views are possible: If the Show all option is activated, all transducer
factors in the directory are displayed. If the Show compatible option is activated, only the
compatible transducer factors in the directory are displayed.
For details on transducer factors refer to chapter "Advanced Measurement Examples".
This softkey is available from firmware version 1.10.
Remote: CORR:TRAN:SEL FACTOR1
Active On/Off
Activates or deactivates the selected transducer factor. A maximum of 10 transducer factors can
be activated at the same time. If an additional transducer factor is activated, its unit compatibility
is checked. If one factor has a different unit than dB, all the other activated factors must have the
unit dB.
This softkey is available from firmware version 1.10.
Remote: CORR:TRAN:SEL <name>
Remote: CORR:TRAN ON
Edit
Opens the Edit Transducer dialog box with the data of the selected factor. During editing, a
transducer factor remains stored in the background until the factor is saved using the Save
Factor softkey.
This softkey is available from firmware version 1.10.
Remote: Comment: CORR:TRAN:COMM <string>
Insert Line
Inserts an empty line above the selected reference value to enter a new reference value. When
entering a new reference value in the line, the ascending order of frequencies must be taken into
consideration, however.
This softkey is available from firmware version 1.10.
Delete Line
Deletes the selected reference value (complete line). The reference values that follow move one
line up.