Status Reporting System R&S FSL
1300.2519.12 5.16 E-11
Overview of the Status Registers
The following figure shows the status registers used by the R&S FSL base unit. The status registers
used by the R&S FSL options are described in separate sections at the end of this chapter.
S
RE
S
TB
PPE
ISTflag
&=logic AND
=logic OR
of all bits
ES E ESR
Error/event
queue
bla
Output
buffer
SRQ
RQS/MSS
ESB
MAV
Power on
User Request
Command Error
Exec uti on Error
Device Dependent Error
Query Error
Request Control
Operation Complete
1
5
14
1
3
1
2
11
10
9
8
7
6
5
4
3
2
1
0
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
7
6
5
4
3
2
1
0
STATus:QUEStionabl e
TRANsducer break
LMARgin
LIMit
CALibration (= UNCAL)
FREQuency
TEMPerature
POW er
7
6
5
4
3
2
1
0
ACPLimit
1
5
1
4
1
3
12
11
10
9
8
7
6
5
4
3
2
1
0
LIMit 8 FAIL
LIMit 7 FAIL
LIMit 6 FAIL
L
IMit 5 FAIL
L
IMit 4 FAIL
LIMit 3 FAIL
LIMit 2 FAIL
LIMit 1 FAIL
STATus:QUEStionable:POWer
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
IF_OVerload
UNDerload
OVERload
STATus:QUEStionable: FREQuency
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
LO UNLocked
OVEN COLD
LMARgin 8 FAIL
LMARgin 7 FAIL
LMARgin 6 FAIL
L
MARgin 5 FAIL
LMARgin 4 FAIL
LMARgin 3 FAIL
LMARgin 2 FAIL
LMARgin 1 FAIL
15
1
4
1
3
12
11
10
9
8
7
6
5
4
3
2
1
0
ST ATus:QUEStionable:LMARgin <1|2>
STATus:OPERation
1
5
1
4
1
3
12
1
1
1
0
9
8
7
6
5
4
3
2
1
0
n
ot used
Scan results available
H
COPy in progress
CALibrating
STATus:QUEStionable:LI Mit <1|2>
-&-
-
&-
-&-
-&-
-&-
-&-
15
14
1
3
1
2
1
1
10
9
8
7
6
5
4
3
2
1
0
STATus:QUEStionable:ACPLimit
n
ot used
not used not used
not used not used
SYNC
1
5
14
13
12
11
1
0
9
8
7
6
5
4
3
2
1
0
n
ot used
CARRier overload
No carrier
SYNC not found
BURSt not found
STATus:QUEStionable:SYNC
EXTernalREFerence
BATTERY LOW
not used
ALT3 … 11 LOWer/UPPer FAIL
ALT2 LOWer FAIL
ALT2 UPPer FAIL
ALT1 LOWer FAIL
A
LT1 UPPer FAIL
A
DJ LOWer FAIL
ADJ UPPer FAIL
Fig. 5–4 Overview of the status registers (base unit)