CALCulate Subsystem R&S FSL
1300.2519.12 6.40 E-11
C
ALCulate<1|2>:LIMit<1...8>:ACPower:ALTernate<1...11>:RESult?
This command queries the result of the limit check for the alternate adjacent channels for
adjacent channel power measurements.
The numeric suffix after ALTernate denotes the alternate channel. The numeric suffixes <1|2>
w
ith CALCulate and <1...8> with LIMit are irrelevant for this command.
If the power measurement of the adjacent channel is switched off, the command produces a
query error.
This command is a query and therefore has no *RST value.
Parameter
The result is returned in the form <result>, <result> where <result> = PASSED | FAILED and
where the first (second) returned value denotes the lower (upper) alternate adjacent channel.
Example
CALC:LIM:ACP:ALT2 30DB, 30DB
Sets the relative limit value for the power in the lower and upper second alternate adjacent
channel to 30 dB below the channel power.
CALC:LIM:ACP:ALT2:ABS –35DBM, –35DBM
Sets the absolute limit value for the power in the lower and upper second alternate adjacent
channel to –35 dBm.
CALC:LIM:ACP ON
Switches on globally the limit check for the channel/adjacent channel measurement.
CALC:LIM:ACP:ALT2:STAT ON
Switches on the limit check for the lower and upper second adjacent channel.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ALT2:RES?
Queries the limit check result in the second alternate adjacent channels.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, CDMA, EVDO, WLAN