R&S FSL CALCulate Subsystem
1300.2519.12 6.47 E-11
CALCulate:LIMit:LOWer Subsystem
The CALCulate:LIMit:LOWer subsystem defines the lower limit line. This subsystem is available from
firmware version 1.10.
Commands of the CALCulate:LIMit:LOWer Subsystem
– CALCulate<1|2>:LIMit<1...8>:LOWer[:DATA]
– CALCulate<1|2>:LIMit<1...8>:LOWer:STATe
– CALCulate<1|2>:LIMit<1...8>:LOWer:OFFSet
– CALCulate<1|2>:LIMit<1...8>:LOWer:MARGin
– CALCulate<1|2>:LIMit<1...8>:LOWer:MODE
– CALCulate<1|2>:LIMit<1...8>:LOWer:SHIFt
– CALCulate<1|2>:LIMit<1...8>:LOWer:SPACing
– CALCulate<1|2>:LIMit<1...8>:LOWer:THReshold
CALCulate<1|2>:LIMit<1...8>:LOWer[:DATA]
This command defines the values for the selected lower limit line.
The number of values for the CONTrol axis and for the corresponding LOWer limit line has to be
identical. Otherwise default values are entered for missing values or not necessary values are
deleted.
The unit must be identical with the unit selected by CALCulate<1|2>:LIMit<1...8>:UNIT.
If no unit is indicated, the unit defined with CALCulate<1|2>:LIMit<1...8>:UNIT is
automatically used.
If the measured values are smaller than the LOWer limit line, the limit check signals errors.
The units DEG, RAD, S, HZ, PCT are not available in the spectrum analyzer mode.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
<numeric_value>,<numeric_value>
Example
CALC:LIM2:LOW –30,–40,–10,–40,–30
Defines 5 lower limit values for limit line 2 in the preset unit.
CALC:LIM2:LOW?
Outputs the lower limit values of limit line 2 separated by a comma.
Characteristics
*RST value: – (LIMit:STATe is set to OFF)
SCPI: conform
Mode
A, ADEMOD, CDMA, EVDO, NF