R&S FSL CALCulate Subsystem
1300.2519.12 6.81 E-11
CALCulate<1|2>:MARKer<1...4>:FUNCtion:TOI:SEARchsignal
This command starts the search of the signals required for the third order intercept
measurement.
The numeric suffix <1|2> are irrelevant for this command.
Note: No new measurement is done. Only the currently available trace selected for the TOI
measurement is used
Parameter
ONCE
Example
CALC:MARK:FUNC:TOI:SEAR ONCE
Executes the search for 2 signals and their intermodulation product at the currently available
trace.
Characteristics
*RST value: –
SCPI: conform
Mode
A
CALCulate<1|2>:MARKer<1...4>:FUNCtion:ZOOM
This command defines the range to be zoomed around marker 1. Marker 1 is activated first, if
necessary.
The subsequent frequency sweep is stopped at the marker position and the frequency of the
signal is counted. This frequency becomes the new center frequency, and the zoomed span is
set. In order to recognize the end of the operation the synchronization to the sweep end should
be activated. This is only possible in single sweep mode.
The numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
Parameter
<numeric_value>
Example
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:ZOOM 1kHz;*WAI
Activates zooming and waits for its end.
CALC:MARK1:FUNC:ZOOM 10
Zooms in around marker 1 by afactor of 10.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–F

Table of Contents

Related product manuals