R&S FSL CALCulate Subsystem
1300.2519.12 6.101 E-11
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:PPEak[:STATe]
This command switches on or off the measurement of the positive peak value.
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
E
xample
CALC:MARK:FUNC:SUMM:PPE ON
Switches on the function.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
A–T
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:PPEak:AVERage:RESult?
This command is used to query the result of the measurement of the averaged positive peak
value. The query is only possible if averaging has been activated previously using
CALCulate<1|2>:MARKer:FUNCtion:SUMMary:AVERage.
A complete sweep with synchronization to sweep end must be performed between switching on
the function and querying the measured value to obtain a correct query result. This is only
possible in single sweep mode.
The numeric suffixes <1|2> are not relevant.
This command is only a query and therefore has no *RST value.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:PPE ON
Switches on the function.
CALC:MARK:FUNC:SUMM:AVER ON
Switches on the calculation of average.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:SUMM:PPE:AVER:RES?
Outputs the result.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T