CALCulate Subsystem R&S FSL
1300.2519.12 6.116 E-11
CALCulate:STATistics:SCALe:Y:UNIT
This command defines the scaling type of the y–axis.
This command is available from firmware version 1.80.
Parameter
PCT | ABS
E
xample
CALC:STAT:SCAL:Y:UNIT PCT
Sets the percentage scale.
Characteristics
RST value: ABS
SCPI: device–specific
Mode
A, EVDO
CALCulate:STATistics:SCALe:Y:UPPer
This command defines the upper limit for the y–axis of the diagram in statistical measurements.
Since probabilities are specified on the y–axis, the entered numeric values are dimensionless.
Parameter
1E–8 to 1.0
Example
CALC:STAT:Y:UPP 0.01
Characteristics
*RST value: 1.0
SCPI: device–specific
Mode
A, CATV, CDMA, EVDO