R&S FSL DIAGnostic Subsystem
1300.2519.12 6.125 E-11
D
IAGnostic<1|2>:SERVice:SFUNction
This command activates a service function which can be selected by indicating the five
parameters: function group number, board number, function number, parameter 1 and
parameter 2 (see service manual). The contents of the parameter string are identical to the code
to be entered in the edit dialog box of manual operation.
The entry of a service function is accepted only if the system password level 1 or level 2 has
been entered previously (SYSTem:PASSword[:CENable]).
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
<string>
Example
DIAG:SERV:SFUN 2.0.2.12.1
Characteristics
*RST value: -
SCPI: device-specific
Mode
all
DIAGnostic<1|2>:SERVice:STESt:RESult?
This command reads the results of the self test out of the instrument. The lines of the result
table are output as string data separated by commas:
"Total Selftest Status: PASSED","Date (dd/mm/yyyy): 09/07/2004
TIME: 16:24:54","Runtime: 00:06","...
The numeric suffixes <1|2> are irrelevant for this command.
Example
DIAG:SERV:STES:RES?
Characteristics
*RST value: -
SCPI: device-specific
Mode
all