Data acquisition and parameters (Sensor menu) R&S NRP
1144.1400.12 4.20 E-2
Sensor menu:
Filter… –
( )
ä
Fig. 4-17 Filter dialog box
A | B | C | D – ( )
ä↔ä
Tabs for selecting the sensor.
Remote Control
In the following commands, [1..4] represents the channel number (1 for channel A, 2 for
channel B, etc).
Auto
[ß ] – ( )
ä ä
Toggles between automatic and manual filter setting.
Auto-filtering is not possible in the Scope mode.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO[?] ON | OFF
Length –
( )
ä ä
For setting the filter length manually. The function can be called only
when autofilter is off. Changes are made immediately when a key is
pressed.
Remote Control
[SENSe[1..4]]:AVERage:COUNt[?] <int_value>
Reset
Initializes the averaging filter with the value currently being measured.
Thus, the display can be updated quickly after a change in level, par-
ticularly in the case of a large filter length.
Remote Control
[SENSe[1..4]]:AVERage:RESet
Auto Once –
( )
ä ä
Determines a “one-off” optimal filter length for the instantaneous
measured value and enters this value in the
Length field. This function
can be called only when autofilter is off.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO[?] ONCE
Auto Config… –
( )
ä ä
Opens a dialog box for configuring the autofilter mode.