Remote Control - Commands R&S NRP
1144.1400.12 6.28 E-3
CALCulate[1..8]:RELative:POWer[:MAGNitude] [?] <float_value>
Has the same direct effect as CALC:REL on the reference value for measured values that share the unit
of a power.
Unit:
DBM | W
Default unit:
DBM
Value range: 1e-18..1e18 W | -150..210 DBM
*RST value: 0 DBM
CALCulate[1..8]:RELative:RATio[:MAGNitude] [?] <float_value>
Has the same direct effect as CALC:REL on the reference value for measured values that share the unit
of a power ratio.
Unit:
DB | PCT
Default unit:
DB
Value range: -200..200 DB | 1e-18..1e22 PCT
*RST value: 0 DB
CALCulate[1..8]:RELative[:MAGNitude]:AUTO[?] OFF | ONCE
Sets the current measured value as the reference value. When used as a setting command, this
command can only be called by the parameter ONCE. OFF will be ignored. OFF is always returned in
response to a query.
*RST value: OFF
Error messages:
-230
"Data corrupt or stale": The current measured value is invalid.
CALCulate[1..8]:RELative:STATe[?] ON | OFF
Determines whether a reference value is used. If this command is called with ON, all results of the
calculate block are divided by the reference value set with
CALCulate[1..8]:RELative[:MAGNitude].
*RST value: OFF