Remote Control - Commands R&S NRP
1144.1400.12 6.34 E-3
DISPlay[:WINDow][1..4]:METer|:ANALog:UPPer[?] <float_value>
Sets the upper limit value of the analog scale.
The value range depends on the current output unit of the measured value of the window.
Unit of
measured
value
W DBM DBUV DPCT DB O
Value range
-9.99999E5
..
9.99999E5
-120
..
90
-13.0103
..
196.897
-100
..
999.99
-200
..
200
-1E20
..
1E20
Unit: Æ Table 6-3 on page 6.27.
*RST value: 30 DBM or 60 DB
DISPlay[:WINDow][1..4]:METer|:ANALog:UPPer:POWer[?]
<float_value>
Sets the upper limit value of the analog scale. This value is used if the unit of the measurement result is
Watt, dBm or dBµV, i.e. the measurement result is a power.
Unit: DBM | W | DBUV
Default unit: DBM
Value range: see DISPlay[:WINDow]:METer|:ANALog:UPPer[?]
*RST value: 30 DBM
DISPlay[:WINDow][1..4]:METer|:ANALog:UPPer:RATio[?]
<float_value>
Sets the upper limit value of the analog scale. This value is used if the unit of the measurement result is
dB, percent or the number 1, i.e. the measurement result is a power ratio.
Unit: DB | DPCT | O
Default unit: DB
Value range: see DISPlay[:WINDow]:METer|:ANALog:UPPer[?]
*RST value: 60 DB
DISPlay[:WINDow][1..4]:NAME[?] <string>
Determines the string to be displayed in the titlebar of the respective window. The reset values depend
on the window (<n> = 1, 2, 3 or 4).
*RST value: ""
Example:
DISP1:NAME "Power In"
DISP2:NAME "Power Out"
The measured values in the display are now visually assigned to the different test points in the test
setup. (Input power in window 1, output power in window 2).