Remote Control - Commands R&S NRP
1144.1400.12 6.46 E-3
OUTPut:RECorder[1..2]:LIMit:LOWer:RATio[?] <float_value>
Specifies the lower power limit of the characteristic for one of the two analog outputs if the associated
calculate block returns a power ratio (unit DB, DPCT or O) as the measured value (ÆFigure 6-5).
Value range: -200 DB .. 200 DB
Unit: DB | DPCT | O
Default unit: DB
*RST value: 0 DB
OUTPut:RECorder[1..2]:LIMit:UPPer[?] <float_value>
Specifies the upper power limit of the characteristic for one of the two analog outputs (ÆFigure 6-5).
The unit corresponds to the output unit of the measured value.
The value range also depends on the current output unit of the measured value in the window:
Unit of
measured
value
W DBM DBUV DPCT DB O
Value range
-9.99999E5
..
9.99999E5
-120
..
90
-13.0103
..
196.897
-100
..
999.99
-200
..
200
-1E20
..
1E20
Unit: Æ Table 6-3 on page 6.27.
*RST value: 20 DBM or 10 DB
OUTPut:RECorder[1..2]:LIMit:UPPer:POWer[?] <float_value>
Specifies the upper power limit of the characteristic for one of the two analog outputs if the associated
calculate block returns a power (unit DBM, W or DBUV) as the measured value
(ÆFigure 6-5).
Value range: -120 DBM to 90 DBM
Unit: DBM | W | DBUV
Default unit: DBM
*RST value: 20 DBM
OUTPut:RECorder[1..2]:LIMit:UPPer:RATio[?] <float_value>
Specifies the upper power limit of the characteristic for one of the two analog outputs if the associated
calculate block returns a power ratio (unit DB, DPCT or O) as the measured value (ÆFigure 6-5).
Value range: -200 DB .. 200 DB
Unit: DB | DPCT | O
Default unit: DB
*RST value: 10 DB