R&S NRP Remote Control - Commands
1144.1400.12 6.95 E-3
SYSTem:INFO? [<string>]
Returns information about the system. <string_value> is used to query a specific information item. If
called without parameters, the command returns all available information in the form of a list of strings
separated by commas. <string_value> can have the following values. No difference is made
between upper-case and lower-case letters:
<string_value>
Meaning
"Manufacturer"
Manufacturer of device
"Type"
Type (R&S NRP)
"Stock Number"
Order No. of R&S NRP
"Serial"
Serial number
"HWVersion"
Hardware version
"SW Build"
Firmware version
"BootloadVer"
Version of boot loader
"KeybCtrlVer"
Version of keyboard controller
"Options"
Installed options
"MACAddr"
MAC address of the R&S NRP if Ethernet option has been installed
"RAMSize"
RAM size
"ROMSize"
ROM size
"CPUClock"
Clock frequency of processor
"CPLDVersion"
Version of CPLD chip
"PICVersion"
Version of PIC-Chip
SYSTem:ERRor?
This query returns error numbers and text from the error/event queue of the R&S NRP. A specific error
number with text is entered in the error/event queue for each error that occurs in the R&S NRP. The
queue functions according to the FIFO principle. The message entered first in the queue is also output
first when a SYST:ERR? query is sent
. Querying clears the message.
Up to 30 errors can be stored in the error/event queue. If further errors occur, the most recent error is
overwritten by the message -350, "Queue overflow".
If the error/event queue is empty, the message 0, "No error" is returned as a response.
The queue is cleared by *CLS. It is not changed by *RST.
For further information on the error/event queue see section Æ"STATus", page 5.1.