R&S NRP Remote Control - Commands
1144.1400.12 6.123 E-3
SENSe[1|2]:CORRection:GAIN2[?]
Has the same effect as SENSe[1..2]:CORRection:OFFSet[?].
With :GAIN2, no unit may be specified. Unit 1 is valid.
SENSe[1|2]:CORRection:LOSS2[?]
Corresponds to the reciprocal of
SENSe[1..2]:CORRection:OFFSet[?]
. With :LOSS2, no unit
may be specified. Unit 1 is valid.
SENSe[1|2]:CORRection:GAIN3[?]
Has the same effect as SENSe[1..2]:CORRection:DCYCle[?].
With :GAIN3, no unit may be specified. Unit 1 is valid.
SENSe[1|2]:CORRection:GAIN4?
Has the same effect as
SENSe[1..2]:CORRection:FDOFfset[:INPut][:MAGNitude]?
SENSe:SWEep:OFFSet:TIME[?]
Is implemented in the R&S NRP and has the same effect as
SENSe:TGATe:OFFSet:TIME[?]
SENSe:SWEep:TIME[?]
Is implemented in the R&S NRP and has the same effect as
SENSe:TGATe:TIME[?]
SYSTem
:COMMunicate:SERial
:HELP:HEADers?
:LOCal
:REMote
:RINTerface
:RWLock
Is not implemented in the R&S NRP.
TRACe:STATe
Is not implemented in the R&S NRP.
TRACe:DATA?
Is not implemented in the R&S NRP in this form. Data of the Scope
mode can only be queried via the high-level commands (FETCh?,
READ?, MEAS?) and CALC:DATA?.
TRACe:UNIT?
The UNIT node is only globally available in the R&S NRP and affects
all measurement results (including Scope data) and the unitless
command parameters of the commands CALC:LIM:LOW,
CALC:LIM:UPP, DISP:MET|:ANAL:LOW, DISP:MET|:ANAL:UPP,
OUTP:REC:LIM:LOW and OUTP:REC:LIM:UPP.
TRIGger[:SEQUence[1]|2]
The R&S NRP does not recognize the:SEQUence header and uses no
TRIGger command globally but only on a sensor-by-sensor basis. If
TRIGger or SENSe commands are to have a global effect, the:ALL
header must be used instead of the numeric suffix.
UNIT:POWer
The R&S NRP recognizes the units W, DBM and DBUV.
UNIT:POWer:RATio
The R&S NRP recognizes the units DB, DPCT and O.
:AC
Agilent recognizes the keyword :AC of some high-level commands
and SENSe
commands:
CONFigure | READ | MEASure | FETCh
[:SCALar][:POWer:AC]
:RELative
:DIFFerence
:DIFFerence:RELative
:RATio
:RATio:RELative
SENSe:POWer:AC:RANGe:AUTO
The R&S NRP also recognizes these commands but
:AVG should be
used instead of
:AC since R&S NRP sensors eventually also measure
DC components and the designation in the R&S NRP is not used in the
sense of SCPI.