R&S NRP SCPI Error Messages
1144.1400.12 9.7 E-2
Error code Error text returned upon queue query
Explanations
54 Sample error
Samples were lost while sampling the signal applied. The measured value can thus be incorrect.
55 Sensor hardware error
A sensor signalled an error in the sensor hardware. If the problem persists, please contact the R&S servicing
center.
56 Filter truncated
With the automatic filter function activated, a sensor tries to set the filter length such that the required accuracy is
obtained. If the measurement time required for this exceeds the value set with
SENSe:AVERage:COUNt:AUTO:MTIMe (max. time), the filter length is shortened such that a measurement does
not last more than the max. time and error message 56 "Filter truncated" is generated.
57 Standard settings incomplete
The R&S NRP can be configured for measuring standardized signals with SYSTem:STANdard:PRESet
<string>. This command sets a number of predefined parameters. If a sensor does not support a specific
parameter (e.g. a measurement mode), execution error message 57 "Standard settings incomplete" is generated.