R&S NRP Measuring average power in defined time interval
1144.1400.12 2.23 E-2
2 Settings to ensure reliable triggering
(¶ MENU)
*
¾ Close the Mode dialog box.
Trigger ( )
* *
¾ Open the Trigger dialog box.
The Timeslot mode provides not only the
Source and Level
parameters described in the section ) Settings for reli-
able triggering but also the
Delay parameter which is used
when the start of timeslot 1 (Frame Start) does not coin-
cide with to the trigger time.
¾ Delay
Define the start of timeslot 1 with reference to the
triggering edge. The value can be positive or nega-
tive.
9 Tip
The Scope display shows physical and delayed trig-
ger times in the form of a triangle so that
Delay can be
interactively adjusted to the signal.
3 Simultaneous measurement in multiple timeslots.
( )
* *
¾ Change to the Windows menu.
¾ Use the rocker switch beside Timeslot # to view the
various timeslot values in the measurement window.
9 Tip
You can open a separate window for each slot and thus
display up to 4 timeslot values simultaneously.