EasyManua.ls Logo

Rohde & Schwarz R&S RTO Series - Page 129

Rohde & Schwarz R&S RTO Series
133 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setting Up the Instrument
R&S
®
RTO
98Getting Started 1316.0833.02 ─ 10
Two connector pins are located on the front panel. The right pin is on ground level. The
left pin supplies a square wave signal with 1 kHz for low frequency probe compensa-
tion.
1. Connect the BNC connector of the probe to input CH1.
2. Connect the probe's ground connector to the right compensation pin, and the tip
with the left pin.
3. Press AUTOSET.
A square wave appears on the display.
4. Adjust the compensation trimmer of the probe to optimum square wave response.
For details, refer to the documentation of your probe.
Adjusting Passive Probes

Table of Contents

Related product manuals