General instrument settings and instrument setup
R&S
®
SGS100A
158User Manual 1173.9105.02 ─ 13
:TEST:ALL:RESult? on page 302
10.12.2 Keyboard tests
Access:
► Select "SGMA-GUI > Instrument Name > Diagnostic Tests > Keyboard Test".
Use this function to check the proper operation of all front panel elements.
If "Keyboard Test" is enabled, all front panel LEDs except the [POWER ON] are
orange.
The exact test procedure is described in the service manual.
SCPI command:
:TEST:KEYBoard[:STATe] on page 302
10.13 External adjustments
The external adjustment is a protected service procedure, that requires "Protection
Level 2" password. For a detailed test procedure, see the R&S SGS service manual.
Access:
► Select "SGMA-GUI > Instrument Name > Service > External Adjustments".
Depending on the installed options, the R&S SGS is equipped with a TCXO or OCXO
reference oscillator. The reference oscillator is factory calibrated to a specific calibra-
tion value. In the external adjustment dialog, you can select a different (custom) cali-
bration value.
External adjustments