Network operation and remote control
R&S
®
SGS100A
194User Manual 1173.9105.02 ─ 13
Figure 12-5: Graphical overview of the status registers hierarchy
OPER = Operation Status Summary Bit
RQS/MSS = Service Request Generation
ESB = Standard Event Status Summary Bit
MAV = Message Available in Output Queue
QUES = Questionable Status Summary Bit
2 = Error- /Event-Queue
1, 0 = not used
Note: This legend explains the abbreviations to the Status Byte Register.
The R&S SGS uses the following status registers:
Status Byte (STB) and Service Request Enable (SRE), see Chapter 12.5.3, "Sta-
tus byte (STB) and service request enable register (SRE)", on page 197.
Status reporting system