Remote control commands
R&S
®
SGS100A
225User Manual 1173.9105.02 ─ 13
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage: Event
13.3
General commands
:LOCK?.........................................................................................................................225
:UNLock........................................................................................................................225
:LOCK? <LockRequestId>
Sends a lock request ID which uniquely identifies the controller to the instrument.
Parameters:
<LockRequestId> Number
0
test query to prove whether the instrument is locked
Controller ID
request lock from the controller with the specified Controller ID
Return values:
<Value> Number
0
request refused; the instrument is already locked to other <Lock
Request Id>, i.e. to another controller
1
request granted
Example:
:LOCK? 12345
Response: 1
:UNL 12345
Usage: Query only
Manual operation: See "Exclusive Access" on page 66
:UNLock <UnlockId>
Unlocks an instrument locked to a controller with Controller ID = <Unlock Id>.
General commands