Remote control commands
R&S
®
SGS100A
229User Manual 1173.9105.02 ─ 13
:CALibration:OEXTension <OExtension>
This SCPI command is used when the instrument is in extension mode.
Enables you to run an internal adjustment only on the extension instrument.
Parameters:
<OExtension> 1 | ON | 0 | OFF
*RST: 0
Example:
:CAL:OEXT ON
activates the mode for calibrating only the extension instrument
:CAL:LEV:MEAS?
calibrates only the extension instrument
:CAL:OEXT OFF
deactivates the mode for calibrating only the extension instru-
ment
:CALibration:ROSCillator:DATA:MODE
<Mode>
Defines whether the factory provided or a custom defined calibration value is used to
adjust the reference oscillator.
Parameters:
<Mode> FACTory | CUSTomer
*RST: FACTory
Example:
:CALibration1:ROSCillator:DATA:MODE CUSTomer
:CALibration1:ROSCillator:DATA 35600
Manual operation: See "Active Adjustment Data" on page 159
:CALibration:ROSCillator[:DATA] <Data>
Sets the calibration value for the custom defined external adjustment.
Parameters:
<Data> integer
Range: 0 to INT_MAX
*RST: 0
Example: See :CALibration:ROSCillator:DATA:MODE on page 229
Manual operation: See "Calibration Value" on page 159