Remote control commands
R&S
®
SGS100A
256User Manual 1173.9105.02 ─ 13
NSRatio
Selects the filter length (averaging factor) according to the crite-
rion that the intrinsic noise of the sensor (2 standard deviations)
does not exceed the specified noise content. You can define the
noise content with command :SENSe<ch>[:POWer]:FILTer:
NSRatio.
Note: To avoid long settling times when the power is low, you
can limit the averaging factor limited with the "timeout" parame-
ter (:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe).
*RST: AUTO
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 247.
Manual operation: See "Filter" on page 122
:SENSe<ch>[:POWer]:FREQuency
<Frequency>
Sets the RF frequency of the signal, if signal source "USER" is selected (:
SENSe<ch>[:POWer]:SOURce).
Parameters:
<Frequency> float
*RST: 1 GHz
Example:
:SENS1:SOUR USER
Selects user-defined source.
:SENS1:FREQ 2.44GHz
Sets the RF frequency of the source which is 2.44 GHz.
Manual operation: See "Frequency" on page 122
:SENSe<ch>[:POWer]:LOGGing:STATe <State>
Activates the recording of the power values, measured by a connected R&S NRP
power sensor.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
:SENS:LOGG:STAT ON
Activates recording of the power measurement of the first sen-
sor.
Manual operation: See "Enable Logging" on page 124
SENSe, READ, INITiate and SLISt subsystems