Remote control commands
R&S
®
SGS100A
259User Manual 1173.9105.02 ─ 13
Manual operation: See "Sensor type and serial number" on page 121
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note: Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
The temperature has varied more than about 5 °C.
The sensor has been replaced.
You want to measure very low power.
Example:
:SENS1:ZERO
Executes zeroing.
Usage: Event
Manual operation: See "Zero" on page 121
13.13 SOURce subsystem
[:SOURce]:OPMode....................................................................................................... 259
[:SOURce]:FREQuency[:CW|FIXed].................................................................................259
[:SOURce]:FREQuency:OFFSet......................................................................................260
[:SOURce]:LOSCillator:SOURce......................................................................................260
[:SOURce]:PATH:COUNt?...............................................................................................261
[:SOURce]:OPMode <OpMode>
Sets the operation mode.
Parameters:
<OpMode> NORMal | BBBYpass
NORMal
normal operation
BBBYpass
Baseband bypass mode
*RST: NORMal
Example: See Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215.
Manual operation: See "Operation Mode" on page 91
[:SOURce]:FREQuency[:CW|FIXed] <Cw>
Sets the RF frequency at the RF output connector of the selected instrument.