Remote control commands
R&S
®
SGS100A
273User Manual 1173.9105.02 ─ 13
Usage: Event
Manual operation: See "Reset Delta Phase Display" on page 93
13.17 SOURce:POWer subsystem
[:SOURce]:POWer:ALC:DSENsitivity............................................................................... 273
[:SOURce]:POWer:ALC[:STATe]...................................................................................... 273
[:SOURce]:POWer:ALC:SONCe......................................................................................273
[:SOURce]:POWer:ATTenuation:RFOFf:MODE................................................................. 274
[:SOURce]:POWer:ATTenuation:SOVer[:OFFSet]..............................................................274
[:SOURce]:POWer:LMODe............................................................................................. 274
[:SOURce]:POWer:POWer..............................................................................................274
[:SOURce]:POWer:SCHaracteristic..................................................................................275
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]..........................................................275
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet................................................................ 276
[:SOURce]:POWer:LIMit[:AMPLitude]...............................................................................276
[:SOURce]:POWer:PEP?................................................................................................ 276
[:SOURce]:POWer:RANGe:LOWer?................................................................................ 277
[:SOURce]:POWer:RANGe:UPPer?.................................................................................277
[:SOURce]:POWer:ALC:DSENsitivity <Sensitivity>
Sets the power detector sensitivity. Used for compatibility reasons only.
Parameters:
<Sensitivity> OFF | LOW | MED | HIGH
*RST: OFF
Manual operation: See "Detector Sensitivity" on page 104
[:SOURce]:POWer:ALC[:STATe] <State>
Activates/deactivates automatic level control.
Parameters:
<State> 1 | OFFTable | OFF | ONTable | AUTO | ON
*RST: ONTable
Example: See Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215.
Manual operation: See "State" on page 104
[:SOURce]:POWer:ALC:SONCe
Briefly activates level control for correction purposes.
Usage:
Event
Manual operation: See "Readjust" on page 101