Remote control commands
R&S
®
SGS100A
276User Manual 1173.9105.02 ─ 13
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet <Offset>
Note: The level offset is also effective for level sweeps!
The command specifies the constant level offset of a downstream attenuator/amplifier.
If a level offset is entered, the level entered with [:SOURce]:POWer:POWer no longer
corresponds to the RF output level.
The following correlation applies:
:POWer = RF output level + POWer:OFFSet.
Entering a level offset does not change the RF output level, but rather the query value
of :POWer.
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.1
*RST: 0
Manual operation: See "Offset" on page 99
[:SOURce]:POWer:LIMit[:AMPLitude] <Amplitude>
Sets the upper limit of the RF signal power.
The value is not affected by an instrument preset and *RST function. This parameter is
influenced only by the factory preset (SYST:FPR) and its factory value is equal to the
upper limit.
Parameters:
<Amplitude> float
Range: -120 to 25
Increment: 0.01
*RST: 25
Default unit: dBm
Example: See Chapter 13.1.4, "Advanced tasks to optimize performance",
on page 218.
Manual operation: See "Limit" on page 101
[:SOURce]:POWer:PEP?
Queries the RF signal's peak envelope power.
Return values:
<PEP> float
Example: see Chapter 13.1.2, "Generating an I/Q modulated signal",
on page 215
Usage: Query only
SOURce:POWer subsystem