EasyManua.ls Logo

Rohde & Schwarz R&S SGS100A - Page 280

Rohde & Schwarz R&S SGS100A
326 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SGS100A
280User Manual 1173.9105.02 ─ 13
Example:
PULM:PER 220 us
the pulse period is 220 us.
Manual operation: See "Pulse Period" on page 127
[:SOURce<hw>]:PULM:POLarity <Polarity>
Sets the polarity of the pulse modulator signal. This command is effective only for an
external modulation signal.
Parameters:
<Polarity> NORMal | INVerted
NORMal
The RF signal is suppressed during the pulse pause.
INVerted
The RF signal is suppressed during the pulse.
*RST: NORMal
Example:
PULM:SOUR EXT
selects the external modulation source.
Example:
PULM:POL INV
selects inverted polarity.
Manual operation: See "Polarity" on page 126
[:SOURce<hw>]:PULM:SOURce <Source>
Selects the source for pulse modulation.
Parameters:
<Source> INTernal | EXTernal
INTernal
The internal pulse generator is used for the pulse modulation.
EXTernal
The signal applied externally via the trigger connector is used for
the pulse modulation.
*RST: INTernal
Manual operation: See "Source" on page 126
[:SOURce<hw>]:PULM:STATe <State>
Activates the pulse modulation.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
PULM:STAT ON
activates pulse modulation.
SOURce:PULM Subsystem