Signal generator settings
R&S
®
SGS100A
99User Manual 1173.9105.02 ─ 13
9.5.1 RF level
Access:
► Select "SGMA-GUI > Instrument Name > Level > RF Level".
This dialog comprises settings like the RF level and level limit.
Level/Level Offset
Sets the RF level at the RF output connector of the selected instrument.
If you set a level offset, it will be indicated in the R&S SGMA-GUI main panel by a
change in the name of this parameter from "Level" to "Level Offset".
Note: The SCPI command [:SOURce]:POWer[:LEVel][:IMMediate][:
AMPLitude] sets the level of the "Level" display, that means the level containing off-
set while [:SOURce]:POWer:POWer sets the level at the RF output connector.
Remote command:
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] on page 275
[:SOURce]:POWer:POWer on page 274
Offset
Sets a level offset.
This value represents the level shift of a downstream instrument, as, for example, an
attenuator or an amplifier, and is indicated in the status bar of the display. It does not
change the level at the RF output.
Level and power-on settings