List and Sweep Mode
R&S
®
SMA100B
171User Manual 1178.3834.02 ─ 05
Settings
Sweep Type ............................................................................................................... 171
State (RF frequency sweep)....................................................................................... 171
State (RF level sweep)................................................................................................171
State (LF frequency sweep)........................................................................................171
Current Frequency...................................................................................................... 172
Current Level...............................................................................................................172
Mode........................................................................................................................... 172
Retrace........................................................................................................................173
Shape..........................................................................................................................173
Spacing....................................................................................................................... 174
Sweep Time ............................................................................................................... 174
Dwell Time ..................................................................................................................174
Trigger Slope...............................................................................................................175
Reset Sweep ..............................................................................................................175
Execute Single Sweep ............................................................................................... 175
Sweep Type
Selects the kind of the frequency sweep.
"Step"
Increases or decreases the frequency at each step.
To determine the frequency intervals, see Spacing.
"Ramp"
Provides a synthesized continuous analog frequency sweep with the
set Sweep Time .
Remote command:
[:SOURce<hw>]:SWEep:GENeration on page 646
State (RF frequency sweep)
Activates RF frequency sweep signal generation.
Note: Active sweep mode deactivates other sweeps or lists and vice versa.
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 591
State (RF level sweep)
Activates RF level sweep signal generation.
Note: Active sweep mode deactivates other sweeps or lists and vice versa.
Remote command:
[:SOURce<hw>]:POWer:MODE on page 633
State (LF frequency sweep)
Status of the LF frequency sweep signal.
You find this parameter in two dialogs:
"LF Frequency Sweep" turns on/off the signal generation.
See Sweep Mode Settings for access.
"Analog Modulation Sources" displays the current state (read-only). If it is turned
on, this dialog provides the varying sweep frequency, see Current Frequency.
Note: Active sweep mode deactivates other sweeps or lists and vice versa.
Sweep Mode Settings